This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Advia Centaur Systems - Br (27.29) Asssays (2017-03-14)

Starting date:
March 14, 2017
Posting date:
April 4, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-62898

Affected Products

A. Advia Centaur System - Br (27.29) Assay

B. Advia Centaur Cp System - Br (27.29) Assay

Reason

Siemens Healthcare Diagnostics has confirmed that the Advia Centaur BR (ca 27.29) assay does not dilute linearly when used with multi-diluent 1 kit lots ending in 2577 and all future lots until further notice. When diluted, some samples over-recover the expected neat value by greater than 120%.

Affected products

A. Advia Centaur System - Br (27.29) Assay

Lot or serial number

All lots

Model or catalog number

3896216

Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

511 Benedict Ave

Tarrytown

10591

UNITED STATES



B. Advia Centaur Cp System - Br (27.29) Assay

Lot or serial number

All lots

Model or catalog number

3896216

Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

511 Benedict Ave

Tarrytown

10591

UNITED STATES