ADVIA Centaur System Calibrator (2016-02-16)
- Starting date:
- February 16, 2016
- Posting date:
- March 4, 2016
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-57398
Affected Products
- A. ADVIA CENTAUR CP SYSTEM-CALIBRATOR
- B. ADVIA CENTAUR SYSTEM-CALIBRATORS
Reason
Siemens Healthcare Diagnostics has confirmed a lot to lot bias with calibrator E kit for the ADVIA Centaur Testosterone assay.
Internal investigation demonstrates:
- an average positive bias of 17% across the assay range with calibrator E kit lots ending in 42 as compared to values generated with the master curve.
- an average negative bias of 7% across the assay range with calibrator E kit lots ending in 43 as compared to values generated with the master curve.
- the positive bias of calibrator E kit lots ending in 42 to the master curve compared to the negative bias of calibrator E kit lots ending in 43 to the master curve is the main driver for the negative bias between values generated with calibrator E kit lots ending in 42 and calibrator E kit lots ending in 43.
The bias of the calibrators to the master curve has been corrected with calibrator E kit lots ending in 44.
The performance of the testosterone assay on the ADVIA Centaur CP is not affected.
Calibrator E is also used for calibration of the ADVIA centaur systems cortisol and progesterone assays, however the performance of these assays is not affected.
Affected products
A. ADVIA CENTAUR CP SYSTEM-CALIBRATOR
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
- 04634452
- 04634762
- 04636889
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
New York
UNITED STATES
B. ADVIA CENTAUR SYSTEM-CALIBRATORS
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
- 04634452
- 04634762
- 04636889
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
New York
UNITED STATES