This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

ADVIA Centaur System Calibrator (2016-02-16)

Starting date:
February 16, 2016
Posting date:
March 4, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-57398

Affected Products

  • A. ADVIA CENTAUR CP SYSTEM-CALIBRATOR
  • B. ADVIA CENTAUR SYSTEM-CALIBRATORS

Reason

Siemens Healthcare Diagnostics has confirmed a lot to lot bias with calibrator E kit for the ADVIA Centaur Testosterone assay.

Internal investigation demonstrates: 

  • an average positive bias of 17% across the assay range with calibrator E kit lots ending in 42 as compared to values generated with the master curve. 
  • an average negative bias of 7% across the assay range with calibrator E kit lots ending in 43 as compared to values generated with the master curve. 
  • the positive bias of calibrator E kit lots ending in 42 to the master curve compared to the negative bias of calibrator E kit lots ending in 43 to the master curve is the main driver for the negative bias between values generated with calibrator E kit lots ending in 42 and calibrator E kit lots ending in 43.

The bias of the calibrators to the master curve has been corrected with calibrator E kit lots ending in 44.

The performance of the testosterone assay on the ADVIA Centaur CP is not affected.

Calibrator E is also used for calibration of the ADVIA centaur systems cortisol and progesterone assays, however the performance of these assays is not affected.

Affected products

A. ADVIA CENTAUR CP SYSTEM-CALIBRATOR

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number
  • 04634452
  • 04634762
  • 04636889
Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

511 Benedict Ave

Tarrytown

10591

New York

UNITED STATES



B. ADVIA CENTAUR SYSTEM-CALIBRATORS

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number
  • 04634452
  • 04634762
  • 04636889
Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

511 Benedict Ave

Tarrytown

10591

New York

UNITED STATES