Health product recall

ADVIA Centaur NT-proBNP (PBNP) & Atellica IM NT-proBNP (PBNP) (2020-03-05)

Starting date:
March 5, 2020
Posting date:
May 8, 2020
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-72989

Last updated: 2020-05-08

Affected Products

  1. ADVIA Centaur NT-proBNP (PBNP)
  2. Atellica IM NT-proBNP (PBNP)

Reason

Siemens Healthcare Diagnostics Inc. confirmed an average method comparison bias of up to 20% with ADVIA Centaur NT-proBNP (PBNP) Assay kits and an average method comparison regression bias of up to 13% with Atellica IM NT-proBNP (PBNP) Assay kits. Siemens' investigation verified that the ADVIA Centaur PBNP and Atellica IM PBNP Assays continue to meet the reference intervals listed in the ADVIA Centaur PBNP and Atellica IM PBNP Instructions for Use (IFU) and concordance (agreement) to the predicate method is greater than or equal to 95%. Siemens is currently investigating root cause of the method comparison bias observed and will send a follow-up communication when further information is available.

Affected products

A. ADVIA Centaur NT-proBNP (PBNP)

Lot or serial number

26878023

Model or catalog number

10699449

Companies
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
New York
UNITED STATES

B. Atellica IM NT-proBNP (PBNP)

Lot or serial number

66426027

Model or catalog number

11200589

Companies
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
New York
UNITED STATES