ADVIA Centaur NT-proBNP (PBNP) & Atellica IM NT-proBNP (PBNP) (2020-03-05)
- Starting date:
- March 5, 2020
- Posting date:
- May 8, 2020
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-72989
Last updated: 2020-05-08
Affected Products
- ADVIA Centaur NT-proBNP (PBNP)
- Atellica IM NT-proBNP (PBNP)
Reason
Siemens Healthcare Diagnostics Inc. confirmed an average method comparison bias of up to 20% with ADVIA Centaur NT-proBNP (PBNP) Assay kits and an average method comparison regression bias of up to 13% with Atellica IM NT-proBNP (PBNP) Assay kits. Siemens' investigation verified that the ADVIA Centaur PBNP and Atellica IM PBNP Assays continue to meet the reference intervals listed in the ADVIA Centaur PBNP and Atellica IM PBNP Instructions for Use (IFU) and concordance (agreement) to the predicate method is greater than or equal to 95%. Siemens is currently investigating root cause of the method comparison bias observed and will send a follow-up communication when further information is available.
Affected products
A. ADVIA Centaur NT-proBNP (PBNP)
Lot or serial number
26878023
Model or catalog number
10699449
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
New York
UNITED STATES
B. Atellica IM NT-proBNP (PBNP)
Lot or serial number
66426027
Model or catalog number
11200589
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
New York
UNITED STATES