Advia Centaur Insulin Assays (2017-04-10)

Starting date:
April 10, 2017
Posting date:
May 2, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-63154

Affected Products

A. Advia Centaur System - Insulin Assay

B. Advia Centaur CP System - Insulin Assay

Reason

Siemens has identified that current in-date lots of Advia Centaur Systems Insulin Assay, recover approximately 40% higher than the world health organization (WHO) 1st IRP 66/304 standardization based on slope values. The slope value was 1.40 on the Advia Centaur and 1.42 on the Advia Centaur CP. Corrected (re-standardized) lots will be available by May 2017.

Affected products

A. Advia Centaur System - Insulin Assay

Lot or serial number

04177173

11090173

11587A59

11606A67

11615A70

11624A74

16546174

61612169

75853169

90517170

Model or catalog number

02230141

04618899

Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

511 Benedict Ave

Tarrytown

10591

UNITED STATES



B. Advia Centaur CP System - Insulin Assay

Lot or serial number

04177173

11090173

11587A59

11606A67

11615A70

11624A74

16546174

61612169

75853169

90517170

Model or catalog number

02230141

04618899

Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

511 Benedict Ave

Tarrytown

10591

UNITED STATES