ADVIA Centaur CP System - Procalcitionin (PCT) Assay and ADVIA Centaur System - Procalcitionin (PCT) Assay (2018-02-13)
- Starting date:
- February 13, 2018
- Posting date:
- March 2, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-66076
Affected products
A. ADVIA CENTAUR CP SYSTEM - PROCALCITONIN (PCT) ASSAY
B. ADVIA CENTAUR SYSTEM - PROCALCITONIN (PCT) ASSAY
Reason
Siemens Healthcare Diagnostics has received multiple customer complaints for invalid calibrations with ADVIA centaur systems B·R·A·H·M·S Procalcitonin reagent kits. This issue affects the ADVIA Centaur/XP/XPT and ADVIA Centaur CP Systems. The 'calibrator ratio' parameter may fall below the acceptable range resulting in an invalid calibration preventing customers from generating procalcitonin results.
Affected products
A. ADVIA CENTAUR CP SYSTEM - PROCALCITONIN (PCT) ASSAY
Lot or serial number
10719038
14888038
37183047
37184047
45947047
97643038
Model or catalog number
10378883
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
UNITED STATES
B. ADVIA CENTAUR SYSTEM - PROCALCITONIN (PCT) ASSAY
Lot or serial number
10719038
14888038
37183047
37184047
45947047
97643038
Model or catalog number
10378883
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
UNITED STATES