Health product recall

ADVIA Centaur CP System - Procalcitionin (PCT) Assay and ADVIA Centaur System - Procalcitionin (PCT) Assay (2018-02-13)

Starting date:
February 13, 2018
Posting date:
March 2, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-66076

Affected products

A. ADVIA CENTAUR CP SYSTEM - PROCALCITONIN (PCT) ASSAY

B. ADVIA CENTAUR SYSTEM - PROCALCITONIN (PCT) ASSAY

Reason

Siemens Healthcare Diagnostics has received multiple customer complaints for invalid calibrations with ADVIA centaur systems B·R·A·H·M·S Procalcitonin reagent kits. This issue affects the ADVIA Centaur/XP/XPT and ADVIA Centaur CP Systems. The 'calibrator ratio' parameter may fall below the acceptable range resulting in an invalid calibration preventing customers from generating procalcitonin results.

Affected products

A. ADVIA CENTAUR CP SYSTEM - PROCALCITONIN (PCT) ASSAY

Lot or serial number

10719038

14888038

37183047

37184047

45947047

97643038

 

Model or catalog number

10378883

Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

511 Benedict Ave

Tarrytown

10591

UNITED STATES



B. ADVIA CENTAUR SYSTEM - PROCALCITONIN (PCT) ASSAY

Lot or serial number

10719038

14888038

37183047

37184047

45947047

97643038

 

Model or catalog number

10378883

Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

511 Benedict Ave

Tarrytown

10591

UNITED STATES