This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
Advia Centaur CP System - Calibrator (field correction)
- Starting date:
- August 10, 2012
- Posting date:
- September 3, 2012
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-15566
Recalled Products
A. Advia Centaur CP System - Calibrator
Reason
Siemens is conducting a field correction for Advia Centaur System Calibrator B Lot 40 and below for use with Advia Centaur Prolactin Assay on the Advia Centaur CP System. Siemens has confirmed analytical sensitivity of greater than the 0.3 ng/ml described in the instructions for use with multiple Advia Centaur Prolactin Reagent Lots and multiple Calibrator B Lots when run on the Advia Centaur CP only.
Affected products
A. Advia Centaur CP System - Calibrator
Lot or serial number
- 79354A30
- 81892B30
- 82815B30
- 83401C30
- 86191A40
- 89717C40
Model or catalog number
652707
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.