Health product recall

Advia Centaur CP System- B-Type Natriuretic Peptide (BNP) Assay (2014-03-24)

Starting date:
March 24, 2014
Posting date:
July 7, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-40393

Recalled Products

Advia Centaur CP System- B-Type Natriuretic Peptide (BNP) Assay

Reason

Siemens Healthcare Diagnostics has received complaints of failed calibrations on the Advia Centaur CP system when using BNP kit lots ending in 172. Siemens has confirmed that the failed calibration is due to an error on the Advia Centaur CP master curve card. The master curve card error results in a low calibration deviation outside the specified calibration evaluation range when using kit lots ending in 172 on the Advia Centaur CP only. An invalid calibration status will prevent BNP results from being generated. The master curve card contained in kit lots ending in 172 for the Advia Centaur and Advia Centaur XP platforms is not impacted and there is no calibration issue when using BNP kit lots ending in 172 on those systems.

Affected products

Advia Centaur CP System- B-Type Natriuretic Peptide (BNP) Assay

Lot or serial number

038172

Model or catalog number

02816138

02816634