This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Advia Centaur, Centaur CP, and Centaur XP Systems - Procaclitonin Assay (PCT) (2014-02-21)

Starting date:
February 21, 2014
Posting date:
April 4, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-38767

Recalled Products

A. Advia Centaur CP System - Procaclitonin (PCT) Assay

B. Advia Centaur XP System -Procalcitonin (PCT) Assay

C. Advia Centaur System - Procalcitonin (PCT) Assay

Reason

Siemens Healthcare Diagnostics received customer feedback regarding the clinical utility of Advia Centaur Brahms Procalcitonin (PCT) quality control level 1. Quality control level 1 currently has a range of plus or minus 50% of the target. Siemens internal investigation confirmed that this range was incorrectly set. Siemens has revised the range of Advia Centaur Brahms Procalcitonin (PCT) quality control level 1 to plus or minus 30% of the target. A review of test data confirmed the assay is capable of achieving a narrower range of plus or minus 30% of the target.

Affected products

A. Advia Centaur CP System - Procaclitonin (PCT) Assay

Lot or serial number

132070674

159460674

181442874

211652874

240972874

279970945

969890674

Model or catalog number

10378884

Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

511 Benedict Ave

Tarrytown

10591

New York

UNITED STATES



B. Advia Centaur XP System -Procalcitonin (PCT) Assay

 

Lot or serial number

132070674

159460674

181442874

211652874

240972874

279970945

969890674

Model or catalog number

10378884

Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

511 Benedict Ave

Tarrytown

10591

New York

UNITED STATES



C. Advia Centaur System - Procalcitonin (PCT) Assay

Lot or serial number

132070674

159460674

181442874

211652874

240972874

279970945

969890674

Model or catalog number

10378884

Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

511 Benedict Ave

Tarrytown

10591

New York

UNITED STATES