This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
Advia Centaur, Centaur CP, and Centaur XP Systems - HIV Ag/Ab Combo (Chiv) Assay (Donor Screen Transplant) (2014-03-28)
- Starting date:
- March 28, 2014
- Posting date:
- April 15, 2014
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-39085
Recalled products
- Advia Centaur Sys-HIV Ag/Ab Combo (Chiv) Assay (Donor Screen Transplant)
- Advia Centaur XP Sys-Hiv Ag/Ab Combo (Chiv) Assay (Donor Screen Transplant)
- Advia Centaur CP Sys-HIV Ag/Ab Combo (Chiv) Assay (Donor Screen Transplant)
Reason
Siemens Healthcare Diagnostics has confirmed that certain samples may show human anti-mouse antibodies (HAMA) interference for the Advia Centaur HIV Ag/Ab combo (CHIV) assay. Heterophilic antibodies in human serum can react with reagent immunoglobulins, interfering with in vitro immunoassays. Patients routinely exposed to animals or animal serum products are prone to this interference and anomalous values may be observed. Additional information may be required for diagnosis.
Affected products
A. Advia Centaur Sys-HIV Ag/Ab Combo (Chiv) Assay (Donor Screen Transplant)
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
- 06520528
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave.
Tarrytown
10591
New York
UNITED STATES
B. Advia Centaur XP Sys-Hiv Ag/Ab Combo (Chiv) Assay (Donor Screen Transplant)
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
- 06520528
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave.
Tarrytown
10591
New York
UNITED STATES
C. Advia Centaur CP Sys-HIV Ag/Ab Combo (Chiv) Assay (Donor Screen Transplant)
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
- 06520528
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave.
Tarrytown
10591
New York
UNITED STATES