Health product recall

Advia Centaur, Centaur CP, and Centaur XP Systems - HIV Ag/Ab Combo (Chiv) Assay (Donor Screen Transplant) (2014-03-28)

Starting date:
March 28, 2014
Posting date:
April 15, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-39085

Recalled products

  1. Advia Centaur Sys-HIV Ag/Ab Combo (Chiv) Assay (Donor Screen Transplant)
  2. Advia Centaur XP Sys-Hiv Ag/Ab Combo (Chiv) Assay (Donor Screen Transplant)
  3. Advia Centaur CP Sys-HIV Ag/Ab Combo (Chiv) Assay (Donor Screen Transplant)

Reason

Siemens Healthcare Diagnostics has confirmed that certain samples may show human anti-mouse antibodies (HAMA) interference for the Advia Centaur HIV Ag/Ab combo (CHIV) assay. Heterophilic antibodies in human serum can react with reagent immunoglobulins, interfering with in vitro immunoassays. Patients routinely exposed to animals or animal serum products are prone to this interference and anomalous values may be observed. Additional information may be required for diagnosis.

Affected products

A. Advia Centaur Sys-HIV Ag/Ab Combo (Chiv) Assay (Donor Screen Transplant)

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number
  • 06520528
Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

511 Benedict Ave.

Tarrytown

10591

New York

UNITED STATES



B. Advia Centaur XP Sys-Hiv Ag/Ab Combo (Chiv) Assay (Donor Screen Transplant)

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number
  • 06520528
Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

511 Benedict Ave.

Tarrytown

10591

New York

UNITED STATES



C. Advia Centaur CP Sys-HIV Ag/Ab Combo (Chiv) Assay (Donor Screen Transplant)

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number
  • 06520528
Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

511 Benedict Ave.

Tarrytown

10591

New York

UNITED STATES