This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Advia Centaur and Centaur XP System Instruments (2014-08-20)

Starting date:
August 20, 2014
Posting date:
September 24, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-41489

Recalled products

  1. Advia Centaur System - Instrument Class 3
  2. Advia Centaur XP System - Instrument Class 2
  3. Advia Centaur XP System - Instrument Class IV
  4. Advia Centaur XP System - Instrument Class III
  5. Advia Centaur System - Instrument Class II
  6. Advia Centaur System - Instrument Class IV

Reason

Manufacturer advises that a specific scenario has been identified that causes patient demographics data (patient name, age, sex) from the previous order received from the Laboratory Information System (LIS) to be merged with the next order. The incorrect patient demographics data will display on the Advia Centaur/Advia Centaur XP user interface, print on instrument-generated reports, and will be transmitted to the LIS with the test result.

Affected products

A. Advia Centaur System - Instrument Class 3

Lot or serial number

All lots.

Model or catalog number
  • 04911855
  • 078-A001-14
Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

511 Benedict Ave.

Tarrytown

10591

New York

UNITED STATES



B. Advia Centaur XP System - Instrument Class 2

Lot or serial number

All lots.

Model or catalog number
  • 078-A010-07
  • 09066916
Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

511 Benedict Ave.

Tarrytown

10591

New York

UNITED STATES



C. Advia Centaur XP System - Instrument Class IV

Lot or serial number

All lots.

Model or catalog number
  • 078-A010-07
  • 09066916
Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

511 Benedict Ave.

Tarrytown

10591

New York

UNITED STATES



D. Advia Centaur XP System - Instrument Class III

Lot or serial number

All lots.

Model or catalog number
  • 078-A010-07
  • 09066916
Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

511 Benedict Ave.

Tarrytown

10591

New York

UNITED STATES



E. Advia Centaur System - Instrument Class II

Lot or serial number

All lots.

Model or catalog number
  • 04911855
  • 078-A001-14
Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

511 Benedict Ave.

Tarrytown

10591

New York

UNITED STATES



F. Advia Centaur System - Instrument Class IV

Lot or serial number

All lots.

Model or catalog number
  • 04911855
  • 078-A001-14
Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

511 Benedict Ave.

Tarrytown

10591

New York

UNITED STATES