This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Advantx, Prestilix, Prestige, and Precision Systems (2013-11-11)

Starting date:
November 11, 2013
Posting date:
January 24, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-37623

Recalled products

  1. Advantx RFX/SFX DS Digital Photspot
  2. Advantx Legacy/Legacy D R/F System
  3. Prestilix 1600x
  4. Prestige VH R/F Remote Controlled Table
  5. Precision 500E System

Reason

It has been identified during several preventive maintenance activities that the 4 outer screws that hold the overhead video monitor suspension to the carriage were found missing since installation of the system. There is a potential that your system has not been installed per specified installation requirements. GE Healthcare is not aware of any fall of the overhead monitor suspension. However, missing screws may increase the risk for fall of the monitor suspension. No injury has been reported due to this issue. Missing screws may increase the risk for the monitor to fall.

Affected products

A. Advantx RFX/SFX DS Digital Photspot

Lot or serial number

Not applicable

Model or catalog number
  • S0912PB
Companies
Manufacturer

GE Medical Systems LLC

3200 North Grandview Blvd.

Waukesha

53188

Wisconsin

UNITED STATES



B. Advantx Legacy/Legacy D R/F System

Lot or serial number

Not applicable

Model or catalog number
  • 2258627
Companies
Manufacturer

GE Medical Systems LLC

3200 North Grandview Blvd.

Waukesha

53188

Wisconsin

UNITED STATES



C. Prestilix 1600x

Lot or serial number

Not applicable

Model or catalog number
  • S13001AX
Companies
Manufacturer

GE Medical Systems LLC

3200 North Grandview Blvd.

Waukesha

53188

Wisconsin

UNITED STATES



D. Prestige VH R/F Remote Controlled Table

Lot or serial number

Not applicable

Model or catalog number
  • S14501BT
Companies
Manufacturer

GE Medical Systems LLC

3200 North Grandview Blvd.

Waukesha

53188

Wisconsin

UNITED STATES



E. Precision 500E System

Lot or serial number

Not applicable

Model or catalog number
  • S0910ZC
Companies
Manufacturer

GE Medical Systems LLC

3200 North Grandview Blvd.

Waukesha

53188

Wisconsin

UNITED STATES