Advantx Legacy and Precision 500D System (2013-09-04)
- Starting date:
- September 4, 2013
- Posting date:
- June 12, 2014
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-39977
Recalled products
- Advantx Legacy System
- Precision 500D system
Reason
There is a potential for the internal bolts on the power assist handle to be loose or to loosen and/or fall out. This issue, if present, can result in a loss of calibration and offset of the power assist handle and/or, under extreme condition, sticking of the handle in the on position. These conditions could cause the system to strike the patient.