This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Advantx Legacy and Precision 500D System (2013-09-04)

Starting date:
September 4, 2013
Posting date:
June 12, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-39977

Recalled products

  1. Advantx Legacy System
  2. Precision 500D system

Reason

There is a potential for the internal bolts on the power assist handle to be loose or to loosen and/or fall out. This issue, if present, can result in a loss of calibration and offset of the power assist handle and/or, under extreme condition, sticking of the handle in the on position. These conditions could cause the system to strike the patient.

Affected products

A. Advantx Legacy System

Lot or serial number

Not applicable

Model or catalog number
  • S0910JB
Companies
Manufacturer

GE Medical Systems LLC

3200 North Grandview Blvd.

Waukesha

53188

Wisconsin

UNITED STATES



B. Precision 500D system

Lot or serial number

Not applicable

Model or catalog number
  • 2403791
Companies
Manufacturer

GE Medical Systems LLC

3200 North Grandview Blvd.

Waukesha

53188

Wisconsin

UNITED STATES