Adult Manual Resuscitator
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
| Affected products | Lot or serial number | Model or catalog number |
|---|---|---|
| Adult Manual Resuscitator | 526796 526804 526797 526805 526817 526782 526802 526815 526807 526811 526800 526806 526809 526814 526816 526798 526808 526790 526810 526818 | CPRM1116F |
Issue
Medline was informed by its contract manufacturer that the Adult Manual Resuscitator with mask, bag reservoir, and filter, CPRM1116F, was misassembled. Instead of the filter pre-attached at the exhalation port, it was connected to the patient port where the mask attaches. If the filter is incorrectly attached and not noticed before patient use, there could be a potential delay in patient resuscitation, which may result in life-threatening consequences, including hypoxia, hypercapnia, organ failure or death.
Recall Start Date: May 9, 2025
Additional information
Details
Medline Industries, Lp
Three Lakes Drive, Northfield, Illinois, United States, 60093