This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

ACTIVA Neurostimulator Charging Systems (2016-11-01)

Starting date:
November 1, 2016
Posting date:
November 18, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-61106

Affected Products

Reason

Medtronic has identified an increased number of complaints from customers involving reports of rechargers that are in an unresponsive error state, where the recharger is non-functional with a blank display screen and is beeping every 5 seconds.  When this error state occurs, the recharger is unable to recharge the neurostimulator until the recharger is reset. If the neurostimulator battery is allowed to become fully depleted, this can lead to loss of therapy and return of associated disease-specific symptoms. If the implanted neurostimulator battery is allowed to remain fully depleted, it may overdischarge, resulting in a permanent reduction in battery capacity and the need to recharge more frequently in the future.

For model 37751 Implantable Neurostimulator (INS) rechargers that are used by patients to recharge their INS devices, notification will be provided to active implanting and managing physicians of both spinal cord stimulation (SCS) and deep brain stimulation (DBS) therapies. All model 37751 rechargers that are included in model 37651, 37754 and 97754 kits may be affected by this issue.

For a subset of patients receiving DBS therapy, in rare instances, a loss of DBS therapy may result in a life threatening injury or death. For example, patients being treated for Parkinson's Disease may experience akinetic crisis, and patients treated for epilepsy may experience status epilepticus. Medtronic has not received any reports of life threatening injury or death associated with this issue.

Affected products

A. ACTIVA RC Neurostimulator - Implantable Neurostimulator Recharger

Lot or serial number

All lots

Model or catalog number

37651

Companies
Manufacturer

Medtronic Inc.

710 Medtronic Parkway NE

Minneapolis

55432

Minnesota

UNITED STATES



B. RestroreSensor Multi-Program Rechargeable Neurostimulator-Charging System 

Lot or serial number

All lots

Model or catalog number

37754

Companies
Manufacturer

Medtronic Inc.

710 Medtronic Parkway NE

Minneapolis

55432

Minnesota

UNITED STATES



C. Charging System

Lot or serial number

All lots

Model or catalog number

97754

Companies
Manufacturer

Medtronic Inc.

710 Medtronic Parkway NE

Minneapolis

55432

Minnesota

UNITED STATES