This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
ACTIVA Neurostimulator Charging Systems (2016-11-01)
- Starting date:
- November 1, 2016
- Posting date:
- November 18, 2016
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-61106
Affected Products
Reason
Medtronic has identified an increased number of complaints from customers involving reports of rechargers that are in an unresponsive error state, where the recharger is non-functional with a blank display screen and is beeping every 5 seconds. When this error state occurs, the recharger is unable to recharge the neurostimulator until the recharger is reset. If the neurostimulator battery is allowed to become fully depleted, this can lead to loss of therapy and return of associated disease-specific symptoms. If the implanted neurostimulator battery is allowed to remain fully depleted, it may overdischarge, resulting in a permanent reduction in battery capacity and the need to recharge more frequently in the future.
For model 37751 Implantable Neurostimulator (INS) rechargers that are used by patients to recharge their INS devices, notification will be provided to active implanting and managing physicians of both spinal cord stimulation (SCS) and deep brain stimulation (DBS) therapies. All model 37751 rechargers that are included in model 37651, 37754 and 97754 kits may be affected by this issue.
For a subset of patients receiving DBS therapy, in rare instances, a loss of DBS therapy may result in a life threatening injury or death. For example, patients being treated for Parkinson's Disease may experience akinetic crisis, and patients treated for epilepsy may experience status epilepticus. Medtronic has not received any reports of life threatening injury or death associated with this issue.
Affected products
A. ACTIVA RC Neurostimulator - Implantable Neurostimulator Recharger
Lot or serial number
All lots
Model or catalog number
37651
Companies
- Manufacturer
-
Medtronic Inc.
710 Medtronic Parkway NE
Minneapolis
55432
Minnesota
UNITED STATES
B. RestroreSensor Multi-Program Rechargeable Neurostimulator-Charging System
Lot or serial number
All lots
Model or catalog number
37754
Companies
- Manufacturer
-
Medtronic Inc.
710 Medtronic Parkway NE
Minneapolis
55432
Minnesota
UNITED STATES
C. Charging System
Lot or serial number
All lots
Model or catalog number
97754
Companies
- Manufacturer
-
Medtronic Inc.
710 Medtronic Parkway NE
Minneapolis
55432
Minnesota
UNITED STATES