Health product recall

Achieva and Intera 1.5T MRI System - Main Unit (2018-03-22)

Starting date:
March 22, 2018
Posting date:
April 13, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-66402

Affected products

Reason

In rare cases, helium gas may escape into the examination room when superconductivity is lost (also known as quenching of the magnet) and at the same time the designed helium ventilation path is impeded. Philips has become aware of one instance where this happened in a Philips system. Investigation found the metal burst disk assembly in the vent path did not perform according to specifications.

Affected products

Achieva 1.5T MRI System- Main Unit

Lot or serial number

All lots.

Model or catalog number

ACHIEVA 1.5T

Companies
Manufacturer

Philips Medical Systems Nederland B.V.

Veenpluis 4-6

Best

5684 PC

NETHERLANDS


Intera 1.5T MRI System-Main Unit

Lot or serial number

All lots.

Model or catalog number

INTERA 1.5T

Companies
Manufacturer

Philips Medical Systems Nederland B.V.

Veenpluis 4-6

Best

5684 PC

NETHERLANDS