Health product recall

ACHIEVA 3.0T MRI, CT, AND XR SYSTEMS (2016-10-12)

Starting date:
October 12, 2016
Posting date:
November 3, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-60802

Affected Products

  1. ACHIEVA 3.0T MRI SYSTEM - MAIN UNIT
  2. ACHIEVA XR SYSTEM - MAIN UNIT
  3. INGENIA 3.0T
  4. INGENIA 3.0T CX

Reason

The thickness and location of the shielding material must be installed according to the shielding requirements provided by Philips. In cases where the requirements are not followed, the distance between the passive shielding rear wall and the magnet could be insufficient. An insufficient distance leads to stronger attraction forces between the magnet and the passive shielding. These stronger forces may lead to shearing of the magnet vibration pads. In a worst case scenario, it may result in displacement of the magnet. Displacement of a magnet can result in image quality issues and mechanical fitting problems with the table. These functional issues affect the clinical usage of the MRI System.

Affected products

A. ACHIEVA 3.0T MRI SYSTEM - MAIN UNIT

Lot or serial number
  • 38109
  • 38195
Model or catalog number
  • ACHIEVA 3.0T
Companies
Manufacturer

Philips Medical Systems Nederland B.V.

Veenpluis 4-6

Best

5684

NETHERLANDS



B. ACHIEVA XR SYSTEM - MAIN UNIT

Lot or serial number
  • 35062
Model or catalog number
  • ACHIEVA XR
Companies
Manufacturer

Philips Medical Systems Nederland B.V.

Veenpluis 4-6

Best

5684

NETHERLANDS



C. INGENIA 3.0T

Lot or serial number
  • 71068
  • 71081
  • 71300
  • 71481
Model or catalog number
  • 3.0T
Companies
Manufacturer

Philips Medical Systems Nederland B.V.

Veenpluis 4-6

Best

5684

NETHERLANDS



D. INGENIA 3.0T CX

Lot or serial number
  • 78012
Model or catalog number
  • 781271
Companies
Manufacturer

Philips Medical Systems Nederland B.V.

Veenpluis 4-6

Best

5684

NETHERLANDS