Achieva 1.5t MRI System-Main Unit
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
---|---|---|
Achieva 1.5t Mri System-Main Unit |
Not applicable. |
ACHIEVA 1.5T |
Issue
Philips has identified an issue where a specific component failure in the gradient coil of the affected 1.5T MR systems listed in this letter may act as a heat source with a potential to produce smoke and/or fire. if the component failure occurs, the user may observe the following:
• repeated scan abort error messages due to gradient amplifier fault detection, described in section 4.
• unusual noise from the system in the examination or technical room
• smoke and/or fire from within the system
Recall start date: October 25, 2023
Additional information
Details
Philips Medical Systems Nederland B.V. |