Access System - Free T3 Assays
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
Unicel DxI 600 Access Immunoassay System - Free T3 Assay |
233968 |
A13422 |
Unicel DxI 800 Access Immunoassay System - Free T3 Assay |
233968 |
A13422 |
Access System - Free T3 Assay |
233968 |
A13422 |
Issue
Beckman Coulter has determined that Access Free T3 reagent lot 233968 included an insufficient concentration of blocking reagent. Patient samples tested with Access Free T3 reagent lot 233968 may demonstrate repeatable falsely elevated results, which are caused by insufficiently blocked interferents.
Recall start date: March 20, 2023