Health product recall

Access IL-6 Calibrators (2015-10-05)

Starting date:
October 5, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-55640

Affected Products

A. Access IL-6 Calibrators
B. Access IL-6 Calibrators
C. Access IL-6 Calibrators

Reason

Beckman Coulter has determined through internal testing that access IL-6 Calibrator Lot 217782 has failed to meet its stability specifications and expiration dating. Access IL-6 Calibrator Lot 217782 may generate falsely elevated patient sample and quality control results across the reportable assay range. Based on internal testing, the customer may experience patient sample and quality control results approximately 15% higher than the corresponding values observed in baseline testing against this calibrator lot.

The issue was discovered through internal testing in April 2015. There has been no report of death or injury related to this issue.

Affected products

A. Access IL-6 Calibrators

Lot or serial number

217782

Model or catalog number

A16370

Companies
Manufacturer
Beckman Coulter
Inc.
250 S. Kraemer Blvd.
92821
UNITED STATES

B. Access IL-6 Calibrators

Companies
Manufacturer
Beckman Coulter
Inc.
250 S. Kraemer Blvd.
92821
UNITED STATES

C. Access IL-6 Calibrators

Lot or serial number

217782

Model or catalog number

A16370

Companies
Manufacturer
Beckman Coulter
Inc.
250 S. Kraemer Blvd.
92821
UNITED STATES