Health product recall

ABX Pentra 400 Clinical Chemistry Systems (2014-09-26)

Starting date:
September 26, 2014
Posting date:
November 26, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-42591

Recalled Products

A) ABX Pentra 400 Clinical Chemistry System - MultiCal

B) ABX Pentra 400 Clinical Chemistry System Controls

Reason

This recall has been initiated due to the potential that, it was observed that creatinine rate blank method calibrator and control results were reporting values lower than target values indicated in the associated ABX Pentra Multical, N control and P control annexes. Horiba has reassigned the target values and updated annexes for the products and lot numbers.

Affected products

A) ABX Pentra 400 Clinical Chemistry System - MultiCal

Lot or serial number

1204103

1304901

1304902

Model or catalog number

A11A01652

Companies
Manufacturer

Horiba ABX, Parc Euromedecine

rue Du Caducee - BP 7290, Cedex 4,

Montpellier

34184

FRANCE



B) ABX Pentra 400 Clinical Chemistry System Controls

Lot or serial number

1300602

1300802

1300803

 

Model or catalog number

A11A01653

A11A01654

Companies
Manufacturer

Horiba ABX, Parc Euromedecine

rue Du Caducee - BP 7290, Cedex 4,

Montpellier

34184

FRANCE