This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
ABL90 Flex Analyzer (2013-10-24)
- Starting date:
- October 24, 2013
- Posting date:
- November 29, 2013
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-36913
Recalled Products
ABL90 Flex Analyzer
Reason
If the ABL90 flex analyzer is tilted it may cause a negative bias on sodium results up to 7% (minus 10mm). This issue has been reported when the analyzer has been installed on a roll stand and tilted (i.e. the support plate is not horizontal).
Affected products
ABL90 Flex Analyzer
Lot or serial number
N/A
Model or catalog number
393-090
Companies
- Manufacturer
-
Radiometer Medical APS
Akandevej 21
Bronshoj
2700
DENMARK