This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

ABL90 Flex Analyzer (2013-10-24)

Starting date:
October 24, 2013
Posting date:
November 29, 2013
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-36913

Recalled Products

ABL90 Flex Analyzer

Reason

If the ABL90 flex analyzer is tilted it may cause a negative bias on sodium results up to 7% (minus 10mm).  This issue has been reported when the analyzer has been installed on a roll stand and tilted (i.e. the support plate is not horizontal).

Affected products

ABL90 Flex Analyzer

Lot or serial number

N/A

Model or catalog number

393-090

Companies
Manufacturer

Radiometer Medical APS

Akandevej 21

Bronshoj

2700

DENMARK