This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

AB SCIEX LC/MS/MS SYSTEM (2016-06-16)

Starting date:
June 16, 2016
Posting date:
July 7, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
Healthcare Professionals, General Public, Hospitals
Identification number:
RA-59258

Affected products

A. API 3200MD LC/MS/MS SYSTEM

B.  3200MD QTRAP LC/MS/MS SYSTEM

C. TRIPLE QUAD 4500MD LC/MS/MS SYSTEM

D. QTRAP 4500MD LC/MS/MS SYSTEM

 

Reason

SCIEX is initiating a field action for the MultiQuant MD Software versions 3.0, 3.0.1 and 3.0.2.  MultiQuant MD software is for use with the API 3200MD LC/MS/MS system, 3200MD QTRAP LC/MS/MS system, triple quad 4500MD LC/MS/MS system and QTRAP LC/MS/MS system. An issue has been identified with MultiQuant MD software where under certain conditions a user can be presented with incorrect quantitative results when using the sum multiple ions feature.

Affected products

A. API 3200MD LC/MS/MS SYSTEM

Lot or serial number

BV20701511

Model or catalog number

5024501

Companies
Manufacturer

AB SCIEX PTE. LTD.,

04-06 MARSILING INDUSTRIAL ESTATE, ROAD 3, BLK 33

739256

SINGAPORE



B.  3200MD QTRAP LC/MS/MS SYSTEM

Lot or serial number

BU20201504

Model or catalog number

5024500

Companies
Manufacturer

AB SCIEX PTE. LTD.,

04-06 MARSILING INDUSTRIAL ESTATE, ROAD 3, BLK 33

739256

SINGAPORE



C. TRIPLE QUAD 4500MD LC/MS/MS SYSTEM

Lot or serial number

BX20951511

Model or catalog number

5031257

Companies
Manufacturer

AB SCIEX PTE. LTD.,

04-06 MARSILING INDUSTRIAL ESTATE, ROAD 3, BLK 33

739256

SINGAPORE



D. QTRAP 4500MD LC/MS/MS SYSTEM

Lot or serial number

BW20201508

Model or catalog number

5031231

Companies
Manufacturer

AB SCIEX PTE. LTD.,

04-06 MARSILING INDUSTRIAL ESTATE, ROAD 3, BLK 33

739256