This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
8mm Tip Cover Accessory
- Starting date:
- August 8, 2012
- Posting date:
- September 17, 2012
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public (GP), Healthcare Professionals, Hospitals
- Identification number:
- RA-15553
Recalled Products
A. 8mm Tip Cover Accessory
Reason
During the manufacturing process, to package each tip cover, each part is placed into the top open end of a rectangular, pre-made pouch which is sealed along the sides and bottom edges. To seal the pouch containing the product, the open end edge is clamped and sealed by applying heat and pressure, creating a closed seal impressed with a thatched pattern. It has been found, with a small number of the pouches manufactured, when creating the closure the sealed edge was not uniform and pleats or creases were formed within the seal area of the pouch. Of those non-uniform seals, in a small percentage of these seals, these pleats or creases may breach the seal barrier integrity.
Affected products
A. 8mm Tip Cover Accessory
Lot or serial number
More than 10 numbers, contact manufacturer
Model or catalog number
400180
Companies
- Manufacturer
-
Intuitive Surgical