Health product recall

840 Ventilator System - Main Unit (2014-06-10)

Starting date:
June 10, 2014
Posting date:
July 10, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-40399

Recalled Products

840 Ventilator System - Main Unit

Reason

Customer reports have included the GUI display screens (both upper and lower) becoming blank during patient use. This event interferes with the clinician's ability to view and/or change ventilator setting or clinician-set patient alarm parameters. The ventilator continues to provide respiratory support during these events. All ventilator alarms remain active and will activate if alarm limits are exceeded.

Affected products

840 Ventilator System - Main Unit

Lot or serial number

Not applicable

Model or catalog number

4-840120DICU-EN
4-840120DICU-FR

Companies
Manufacturer
Covidien LLC
15 Hampshire Street
Mansfield
02048
Massachusetts
UNITED STATES