This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

A-642100 Fixed Offset Adapter (2013-10-25)

Starting date:
October 25, 2013
Posting date:
January 24, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-37537

Recalled products

  1. A-642100 Fixed Offset Adapter

Reason

Company has discovered that the A-642100 offset adapter can develop fatigue cracks in normal use over an extended period of time and ultimately separate completely. Despite the fact that two incidents is not a high failure percentage, the failure mode at hand gives reason to act on the matter to prevent potential patient injury.

Affected products

A. A-642100 Fixed Offset Adapter

Lot or serial number

All lots

Model or catalog number
  • A-642100
Companies
Manufacturer

Ossur HF

Grjothals 5

Reykjavik

110

ICELAND