Health product recall

5mL Reaction Vessels used with the m2000sp Instrument (2014-07-09)

Starting date:
July 9, 2014
Posting date:
August 8, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-40947

Recalled products

  1. 5mL Reaction Vessels used with the m2000sp Instrument

Reason

During run setup on a M2000SP instrument a customer found two 5 ml reaction vessels, list number 04J71-20, with missing rims that fell through the 1ml subsystem carrier. Some of the 5 ml reaction vessels (RV) in the affected lots may have an insufficient rim.  A  RV with an insufficient rim may not be held securely in the 1ml subsystem carrier.  This may cause the RV to be seated incorrectly in the carrier or drop through it.

Affected products

A. 5mL Reaction Vessels used with the m2000sp Instrument

Lot or serial number
  • 56944001
  • 57026001
  • 57059001
  • 57673001
Model or catalog number
  • 04J71-20