Health product recall

3M Durapore Advanced Surgical Tape (2019-11-24)

Starting date:
November 24, 2019
Posting date:
December 20, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-71931


Last updated:

Affected Products

3M Durapore Advanced Surgical Tape

Reason

3M Durapore Advanced Surgical Tape is being recalled due to unanticipated product performance in clinical settings with high humidity conditions and/or when the tape is exposed to normal and/or excessive amounts of oral secretions.  This use environment may cause tape bond failure, leading the tape to not perform as intended.  Further, the tape fibers may fray, enabling the fibers to be accidentally ingested or aspirated.  While the product meets all safety and efficacy criteria, long term exposure of 3M Durapore Advanced Surgical Tape in high moisture clinical applications has not fully met all of 3M's performance expectations.

Affected products

3M Durapore Advanced Surgical Tape

Lot or serial number

All codes.

Model or catalog number

1590-0

1590-1

1590-2

1590-3

15905-1

Companies
Manufacturer

3M Deutschland GmbH Trading as "Health Care Business"

Carl-Schurz-STR.1

Neuss

41453

GERMANY