20/30 Priority Pack Accessory Kit With Copilot and 20/30 Indeflator Inflation Device
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
20/30 Priority Pack Accessory Kit With Copilot |
More than 10 numbers, contact manufacturer. |
1003327 |
20/30 Indeflator Inflation Device |
More than 10 numbers, contact manufacturer. |
1000184 |
Issue
This action is being taken due to an increased potential for a leak in the indeflator at the hose rotating luer or at the guage. Analysis indicates an estimated rate of occurrence of leaks ranging from 0.3% to 2.3% across the impacted lots. While no adverse patient effects have been reported with this issue, potential risks include air embolism.
Recall Start Date: November 6, 2024
Additional information
Details
Abbott Vascular
3200 Lakeside Drive, Santa Clara, California, United States, 95054-2807