RAPIDPoint 500 System (2018-10-02)

Starting date:
October 2, 2019
Posting date:
November 29, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-71728

Last updated: 2019-12-02

Affected Products

RAPIDPoint 500 System

Reason

Siemens has identified an issue with the optional custom demographics software feature on RAPIDPoint 500 analyzers with software versions V2.2, V2.3 or V2.3.1. This optional feature enables you to define up to 10 free text fields. The issue impacts the custom demographics fields sent to the lis and viewed/printed from the data recall screen. These fields may get mixed up with custom demographics from another patient. The sample printout obtained at the time the sample was run is not impacted.

Affected products

RAPIDPoint 500 System

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

10492730

Companies
Manufacturer
Siemens Healthcare Diagnostics Inc.