Oragene DNA collection device (2019-06-28)
- Starting date:
- June 28, 2019
- Posting date:
- July 19, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-70467
Last updated: 2019-07-19
Affected Products
Oragene DNA collection device
Reason
During an internal investigation, it was discovered that the ink on the Oragene DNA collection device tube label from the affected lots could potentially smudge when handled by firmly rubbing on the label. If not properly handled, information on the label may deteriorate over time and impact sample traceability.
Affected products
Oragene DNA collection device
Lot or serial number
WE21
WE24
Model or catalog number
OG-600
OG-610
Companies
- Manufacturer
-
DNA Genotek INC.
3000 - 500 PALLADIUM DRIVE
Ottawa
K2V 1C2
Ontario
CANADA