We have found an insufficient anti-reflective coating on an optical component, which can under certain circumstances lead to an additional axis length signal. If the correct axis length signal is weaker, an incorrect axis length value can be displayed…
RecallHealth product recall | 2023-08-24
During post-market surveillance, we have found that the IOL calculator printout often does not accurately reflect the alignment axis and incision position when planning toric IOLs. The information displayed directly in the software is correct.
Recall…