Mahurkar Acute High Pressure Triple Lumen Catheter Kit Curved Extensions
During manufacturing related testing, the catheter center lumen was found to have an occlusion in the tip of the catheter. The source of the occlusion is excessive mdx, a silicone-based lubricant which coats the catheter tip.
Recall Start Date: June…
RecallHealth product recall | 2023-07-07
Surgical Gut - Plain and Absorbable Sutures
Medtronic was made aware of a supplier calibration issue during the contract sterilization process that may lead to unsterile product.
Recall start date: March 1, 2023
RecallHealth product recall | 2023-03-17
Tri-Staple™ 2.0 Black Intelligent Reload
Specific lots have the potential for a broken sled vane component. A broken sled vane may cause the reload to misfire leading to non-functional staple line closure, transecting tissue without forming staples, and tissue hang-up. These conditions may…
RecallHealth product recall | 2023-02-08