Dräger has become aware of one case in which an obstructed Breathing System Filter Safestar 55 was used on a patient during anesthesia. The patient reportedly became hypoxic and had to be reanimated. If an obstructed filter is used on a patient,…
RecallHealth product recall | 2022-05-10
Bias flow selection for NO system usage is missing from the user interface of Fabian HFOi Ventilators following incorrect selection of device configuration during a software update. This is caused by human error and there is no defect in the device or…
RecallHealth product recall | 2022-04-12
Ventilation stops during clinical use, with waveforms and parameters frozen (not updated). a continuous audible/visual high priority alarm for technical failure 305 is generated. ventilation is suspended until unit is rebooted/replaced.
Recall start…
RecallHealth product recall | 2021-12-31
This voluntary medical device correction is being taken following the investigation of six customer reports stating that a total of four ventilators became inoperable during use. One of these reports indicated that, following the ventilator becoming…