Search
Type
Audience
Category
Issue
Last updated
Recall class
Please note: using the search filters may not provide accurate search results for records prior to November 2021. When searching for older records, please ensure that you use the include archive checkbox.
Displaying 1 - 15 of 3041 items.
Specific lots have the potential for a broken sled vane component. A broken sled vane may cause the reload to misfire leading to non-functional staple line closure, transecting tissue without forming staples, and tissue hang-up. These conditions may…
RecallHealth product recall | 2023-02-08
In the course of internal quality controls, Draeger has noticed that some of their flex ventilation hoses have contaminations. Due to a deviation in the manufacturing process, components of the hose material remained as particles inside the hoses.…
RecallHealth product recall | 2023-02-08
Baxter Corportation is issuing a recall for certain lots of the Minicap. These devices are packaged in foil pouches, which may have been incorrectly sealed, i.e., the pouches may have open or weak seals.
Recall start date: February 01, 2023
RecallHealth product recall | 2023-02-08
Momcozy wearable breast pumps were sold in the Canadian market without the proper authorization (i.e., medical device licence [MDL]).
Recall start date: January 28, 2023
RecallHealth product recall | 2023-02-08
Convatec is conducting a medical device recall for a specific lot of Aquacel® Ag foam adhesive 10cm x 10cm due to the visual contamination in the form of brown spots present on dressings.
An internal investigation determined that the likely root cause…
RecallHealth product recall | 2023-02-07
During the investigation process of customer complaints, Lina has become aware of a safety issue in relation to the use of an obsolete version of an accompanying document. The use of an obsolete customer quick reference guide resulted in the incorrect…
RecallHealth product recall | 2023-02-07
Medline 3" walker wheels with glide caps can potentially come with a loose acorn nut that attaches the wheel to the leg.
Health consequences include the chances of the wheel becoming loose and disconnecting resulting in an unstable condition of the…
RecallHealth product recall | 2023-02-07
Hologic has received a single report of a c-arm performing a significant unexpected movement during non-patient related use of the system. The investigation determined this event was due to a hardware component failure.
Recall start date: January 16,…
RecallHealth product recall | 2023-02-07
Through ongoing internal testing, Baxter Corporation has observed isolated incident of over infusion on the Novum IQ LVP infusion pump during test runs at 0.1 ml/hr after 40 hours of continuous infusion within the 96-hour set change interval.
Baxter…
RecallHealth product recall | 2023-02-06
Product was identified as potentially non-sterile as it received an underdose of gamma irradiation during the sterilization process. As the sterilization dose was not achieved, the sterility assurance level cannot be guaranteed.
Recall start date:…
RecallHealth product recall | 2023-02-01
Non-licensed inventory was sent to Canada via web shop orders starting in February of 2022.
Recall start date: December 12, 2022
RecallHealth product recall | 2023-01-31
During internal studies with the Elecsys® Troponin T hs (high sensitive) and Elecsys® Troponin T hs Stat assays, it was confirmed that for certain k2/k3 EDTA primary tubes, TnT hs results are elevated compared to serum samples when measured…
RecallHealth product recall | 2023-01-31
Beckman Coulter has determined that the Access 2 wash arm lead screw nut band is susceptible to breakage. An Access 2 Immunoassay system wash arm may not move as intended if the nut band is broken. A broken nut band may increase the risk of erroneous…
RecallHealth product recall | 2023-01-31
Summit Medical received a customer complaint that Nasal Septal Button, 5cm, SP-78105 parts received from lot# 219349 were in unsealed pouched. The product was sterilized, but if this sterile barrier is not sealed, there is a possibility of product…
RecallHealth product recall | 2023-01-26
There is a potential malfunction with the side-port of the Triever 24. The side-port may potentially leak or dislodge during device preparation or during actual use. The side-port of the triever24 catheter is used in combination with a 60cc large bore…
RecallHealth product recall | 2023-01-26