Olympus has received complaints that the grasping portion of FG-51D forceps can be difficult to open and close. Olympus reviewed approximately 200 complaints globally over the last five years. Subsequent complaint investigation has confirmed that the FG-…
RecallHealth product recall | 2022-04-11
If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the x-ray tube is no longer sufficiently cooled and the system will display the message "tube hot, have a break". Several minutes later the…
RecallHealth product recall | 2022-03-14
On 10 November 2021, a Corkscrew Needle Electrode dislodged from the wire and hub; and had to be removed from the patient's scalp by making a small incision (NCR 21-081). Later on, NCR 21-087 was received in which the needle had to be removed from the…
RecallHealth product recall | 2022-01-04
Baxter Corporation has internally identified a scenario on the Novum IQ Large Volume Infusion Pump (lVP) where 1) an unintended second bolus or 2) an unintended second auto-programming order update may be delivered to the patient. Baxter Corporation has…
RecallHealth product recall | 2022-01-04
Philips has discovered a problem with the flat detector cooling system. Due to a leak in the flat detector cooling system, cooling liquid may leak outside the drip tray of the laird chiller. If this occurs, the liquid may drip into electrical components…
RecallHealth product recall | 2021-12-13
Loctite may be missing on the stop-screw which holds the hanger-bar hook in place. This is the hook which connects to the distance strap.
Recall start date: 2021-10-18
RecallHealth product recall | 2021-11-12
leica microsystems received one complaint (May 2021) pertaining to the M220 F12 surgical microscope system stating that the optics carrier fell down during the surgical procedure. The issue was found to be related to a loose screw in the…
RecallHealth product recall | 2021-11-09
This correction decision has been made because the transit bracket may potentially fail in certain instances.
Initial recall date: 2021-09-08
RecallHealth product recall | 2021-10-29
The Virtutrax device is not able to be tightened onto an instrument to be used.
Recall date: 2021-10-15
RecallHealth product recall | 2021-10-29