Surestrips Covid-19 Antigen Tests are not authorized for sale in Canada as they were re-packaged into single tests and re-labelled without authorization by the manufacturers listed on the label (i.e. Artron Laboratories Inc.).
Recall start date…
RecallHealth product recall | 2022-05-30
Mold was found in lot Y591676 of the product. Two species of mold were identified that could pose a potential health risk to users.
Recall start date: May 3, 2022
RecallHealth product recall | 2022-05-30
An end user site attempted to perform QC on a single test using multiple QC kits. The QC module was not designed to support testing of multiple QC kits for a single test, and it is required that all QC samples used to validate a test are contained in a…
RecallHealth product recall | 2022-05-30
Surestrips Rapid Response Covid-19 Antigen Rapid Test Device are not authorized for sale in Canada as they were re-packaged into single tests and re-labelled without authorization by the manufacturers listed on the label (i.e.…
RecallHealth product recall | 2022-05-30
After initiating testing by a third-party lab, it was discovered that the tested lots did not meet the astm d6319 standard for medical gloves. Initial testing was received on March 4, 2022 and communicated to the original manufacturer. Confirmatory third…
RecallHealth product recall | 2022-05-20
The recall activities are being conducted due to several temperature excursions on inbound shipment to Canada. According, to the assessment by the manufacturer, Anteis S.A, no data is available for Belotero products after repeated freezing and…
RecallHealth product recall | 2022-05-16
Potential for blood leak to occur with the anti-reflux valve upon treatment completion.
Recall start date: May 5, 2022.
RecallHealth product recall | 2022-05-16
The Carescape Central Station (CSCS) V2 can shut down due to a potential power supply component failure. This can lead to loss of patient monitoring at the central station. Patient monitoring at the bedside is not affected.
Recall start date: May 10,…
RecallHealth product recall | 2022-05-16
During continuous post market surveillance, Draeger gathered information and market feedback indicating that there is a need to provide additional information in the instructions for use regarding the potential of the user to incorrectly position the…
RecallHealth product recall | 2022-05-10
During global post market surveillance activities cases have been reported that the sampling pump 6873493 shows a changing characteristic with increasing operating time, which can lead to the deactivation of the gas measurement during operation. The…
RecallHealth product recall | 2022-05-10
Since Nov. 2020, Arjo has become aware of 2 customer complaints involving overheating of metal part of the device located underneath its frame. The part involved is referred to as the e-bay box. No adverse health consequences (including temporary or…
RecallHealth product recall | 2022-05-10
The STO2 values may be inaccurately low when using either the fore-sight elite tissue oximeter module (Model HEMFSM10) or the fore-sight elite absolute tissue oximeter monitor (Model 01-06-3000) with the fore-sight elite large sensor (…
RecallHealth product recall | 2022-05-09
Customer complaint on over-inflation (abnormal inflation issue) occurring during the product installation, Preparation to initiate therapy or product use. In none of those events, any injury or health consequences were reported. the products in the…
RecallHealth product recall | 2022-05-09
Reports of products having torn corrugated breathing tube.
- Should a tear be detected before use, a delay may be experienced while another product is located.
- Should a tear be detected during use, this may result in a leak and…
RecallHealth product recall | 2022-05-09
A software update for carelink SmartSync™ device managers (SmartSync) will address a telemetry error that may occur with Medtronic Cobalt™ and Crome™ Implantable Cardioverter Defibrillators (ICDs), and Cardiac Resynchronization Therapy Defibrillators (…
RecallHealth product recall | 2022-05-09