This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Zoll Onestep Complete Electrode (2017-08-08)

Starting date:
August 8, 2017
Posting date:
September 8, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-64442

Affected Products

  1. Zoll Onestep Complete Electrodes (Case)
  2. Zoll Onestep Complete Electrode

Reason

A portion (640 out of 3400) of the IDchips in the electrodes for Lot 2517A were inadvertently programmed as pediatric electrodes, instead of adult electrodes.

Affected products

  1. Zoll Onestep Complete Electrodes (Case)

Lot or serial number

2517A

Model or catalog number

8900-0214

Companies
Manufacturer

Zoll Medical Corporation

269 Mill Road

Chelmsford

01824

Massachusetts

UNITED STATES



B. Zoll Onestep Complete Electrode

Lot or serial number

2517A

Model or catalog number

8900-0214

Companies
Manufacturer

Zoll Medical Corporation

269 Mill Road

Chelmsford

01824

Massachusetts

UNITED STATES