Zippie Voyage Early Intervention Stroller (2021-07-23)
- Starting date:
- July 23, 2021
- Posting date:
- August 6, 2021
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-76195
Last updated:
Affected Products
Zippie Voyage Early Intervention Stroller
Reason
On April 2021 an injury was reported in the US where a child suffered a facial laceration, when her Zippie Voyage Stroller seating system detached due to inadequate latching to the stroller base locking mechanism by the child’s caregiver. 13 injuries have been reported in the US. The manufacturer has made the decision for a field correction to implement a safety strap (tether) that when used, will prevent the device seating system from falling to the ground.
Affected products
Zippie Voyage Early Intervention Stroller
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
EIZ14
EIZ16
EIZ18
Companies
- Manufacturer
-
Sunrise Medical (US) LLC
2842 Business Park Avenue
Fresno
93727
California
UNITED STATES