This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Zimmer Trabecular Metal Reverse Shoulder System Instrumentation Liner Impactor

Starting date:
July 19, 2012
Posting date:
September 3, 2012
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-23807

Recalled Products

  1. Zimmer Trabecular metal Reverse Shoulder System Instrumentation Liner Impactor 40mm
  2. Zimmer Trabecular Metal Reverse Shoulder System Instrumentation Liner Impactor 40mm 65 degree Neck Angle
  3. Zimmer Trabecular Metal Reverse Shoulder System Instrumentation Liner Impactor 36mm
  4. Zimmer Trabecular Metal Reverse Shoulder System Instrumentation Liner Impactor 36mm 65 degree Neck Angle

Reason

Lot specific recall due to reports of alignment peg fractures occurring in approximately 0.4% of surgeries. The recalled lots were manufactured from 455 stainless steel, which is the material that all of the reported fractured instruments were manufactured from. Due to the notch sensitivity of the 455 stainless steel, the peg on the Impactor manufactured with this material has a propensity for fracture. The remainder of lots that are not affected by this recall were made with 13-8 stainless steel.

Affected products

A. Zimmer Trabecular metal Reverse Shoulder System Instrumentation Liner Impactor 40mm

Lot or serial number

6109783

Model or catalog number

00-4309-028-00

Companies
Manufacturer

Zimmer Inc.

Zimmer of Canada Ltd.



B. Zimmer Trabecular Metal Reverse Shoulder System Instrumentation Liner Impactor 40mm 65 degree Neck Angle

Lot or serial number

60936815, 61051017

Model or catalog number

00-4309-028-01

Companies
Manufacturer

Zimmer Inc.

Zimmer of Canada Ltd.



C. Zimmer Trabecular Metal Reverse Shoulder System Instrumentation Liner Impactor 36mm

Lot or serial number

61004027

Model or catalog number

00-4309-029-00

Companies
Manufacturer

Zimmer Inc.

Zimmer of Canada Ltd.



D. Zimmer Trabecular Metal Reverse Shoulder System Instrumentation Liner Impactor 36mm 65 degree Neck Angle

Lot or serial number

60989817, 61011689, 61024161

Model or catalog number

00-4309-029-01

Companies
Manufacturer

Zimmer Inc.

Zimmer of Canada Ltd.