Zeus® Infinity® Empowered
Brand(s)
Last updated
Summary
Product
Zeus® Infinity® Empowered
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products | Lot or serial number | Model or catalog number |
---|---|---|
Zeus® Infinity® Empowered | More than 10 numbers, contact manufacturer. | MK04000 |
Issue
In the course of service activities, it was found that CO2 readings were slightly increased. In some cases, the ISO accuracy was no longer achieved. The tolerance check is intended to ensure that the zeus ie will measure all gases within the specified accuracy until the next annual check.
Draeger has updated the instructions for use of the Zeus® ie anesthesia device regarding the test tolerance of the CO2 measurement.
Recall start date: December 23, 2024
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Companies
Draegerwerk Ag & Co. Kgaa
Moislinger Allee 53-55, Lubeck, Germany, 23542
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-76804
Get notified
Receive notifications for new and updated recalls and alerts by category.