Zeus® Infinity® Empowered
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
Zeus® Infinity® Empowered |
More than 10 numbers, contact manufacturer. |
MK04000 |
Issue
In the course of service activities, it was found that CO2 readings were slightly increased. In some cases, the ISO accuracy was no longer achieved. The tolerance check is intended to ensure that the zeus ie will measure all gases within the specified accuracy until the next annual check.
Draeger has updated the instructions for use of the Zeus® ie anesthesia device regarding the test tolerance of the CO2 measurement.
Recall start date: December 23, 2024
Additional information