Zeus IE Anesthesia Workstation (2019-03-11)
- Starting date:
- March 11, 2019
- Posting date:
- April 5, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-69484
Last updated: 2019-04-05
Affected Products
Zeus IE Anesthesia Workstation
Reason
Cases have been reported where the gas measurement of Zeus IE Anesthesia workstations became inactive. Draeger analysis revealed that the deactivations were caused by a too strict supervision of the sample gas pumps which are used in the gas measurement modules. As the characteristic of the pump changes with their usage time, the supervision system deactivates the gas measurement and a corresponding alarm message appears. Consequently, the probability of a gas measurement deactivation is correlated with the total usage time of the sample gas pumps. Long-term tests clearly indicate that the gas measurement is reliable for approximately 7 months assuming an average runtime of 84 hours per week. From that point on the probability of a deactivation increases beyond initial estimations.
Affected products
Zeus IE Anesthesia Workstation
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
- MK04000
Companies
- Manufacturer
-
Draegerwerk AG & Co. KGaA
Moislinger Allee 53-55
Lübeck
23542
GERMANY