This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

ZEUS IE ANESTHESIA WORKSTATION (2015-12-31)

Starting date:
December 31, 2015
Posting date:
April 22, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-57990

Affected products 

A. ZEUS IE ANESTHESIA WORKSTATION

Reason

Firm became aware of an event in which a 38mm pole broke due to excessive load. It was found that the Draeger Infinity Acute Care System was mounted on the optional 38mm pole in a way not validated by Draeger and the maximum load of 13 kg was exceeded. If the weight of the equipment mounted on the 38mm pole exceeds 13kg, it could potentially bend, break or fall on a patient, user or bystander.

Affected products

A. ZEUS IE ANESTHESIA WORKSTATION

Lot or serial number

More than 100 numbers, contact manufacturer.  

Model or catalog number

MK04000

Companies
Manufacturer

DRAEGERWERK AG & CO

MOISLINGER ALLEE 53-55

LUBECK

23542

GERMANY