- Starting date:
- September 15, 2016
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-60450
Affected Products
Reason
Suspension cable can break if user excessively twists the cable in the same direction in each procedure, and prevents the cable from unwinding between procedures.
Affected products
ZERO-GRAVITY SUSPENDED RADIATION PROTECTION SYSTEM
Lot or serial number
All lots.
Model or catalog number
- ZG48
- ZGCH
- ZGCH48
- ZGCM-48
- ZGCM-48-SPECIAL
- ZGCM-66
- ZGCM-HSA
- ZGCSP
- ZGHSA
- ZGM-6-5ROUNDH
- ZGM-6.5H
- ZGM-6DEMO
Companies
- Manufacturer
-
RAWCAR GROUP LLC
CONTOUR FABRI. INC., CFI MEDICAL SOLUTIONS
14241 FENTON ROAD
FENTON
48430
Michigan
UNITED STATES