Zenition 70 and 50
Last updated
Summary
Product
Zenition 70 and 50
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products | Lot or serial number | Model or catalog number |
---|---|---|
Zenition 70 | Not applicable. | 718133 |
Zenition 50 | Not applicable. | 718096 |
Issue
During the system start up or during an ongoing clinical procedure, due to fluctuations in the hospital main power, the fuse can blow thereby causing loss of power to the generator and height positioning module. This results in a system not being available for clinical use. As the system is not available for clinical use, it can result in a delay or abortion of the clinical procedure.
Recall Start Date: September 30, 2024
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies
Philips Medical Systems Nederland B.V.
Veenpluis 6, Best, Netherlands, 5684 PC.
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-76314