Health product recall

Zenition 70 and 50

Last updated

Summary

Product
Zenition 70 and 50
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

Zenition 70

Not applicable.

718133

Zenition 50

Not applicable.

718096

Issue

During the system start up or during an ongoing clinical procedure, due to fluctuations in the hospital main power, the fuse can blow thereby causing loss of power to the generator and height positioning module. This results in a system not being available for clinical use. As the system is not available for clinical use, it can result in a delay or abortion of the clinical procedure.

Recall Start Date: September 30, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies

Philips Medical Systems Nederland B.V.

Veenpluis 6, Best, Netherlands, 5684 PC.

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-76314

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe