Health product recall

Ysio X.pree (2021-03-12)



Starting date:
March 12, 2021
Posting date:
April 1, 2021
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-75277


Last updated:

Affected Products 

Ysio X.pree

Reason

Siemens has received a complaint about an image assigned to a different patient during image recovery process. In the unlikely event of an interrupted connection to a max detector during an initiated image recovery, the previously acquired image (possibly belonging to the previous patient) might be displayed to the user instead of the currently acquired image.

Affected products

Ysio X.pree

Lot or serial number
  • VA10C
Model or catalog number
  • 11107464
Companies
Manufacturer

Siemens Healthcare GMBH

Henkestr. 127

Erlangen

91052

GERMANY