Ysio Radiography X-Ray System - Full Motorized (2020-02-10)
- Starting date:
- February 10, 2020
- Posting date:
- February 28, 2020
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Laceration Hazard
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-72435
Last updated: 2020-03-06
Affected Products
Ysio Radiography X-Ray System - Full Motorized
Reason
Siemens is notifying users of Ysio Fully Automated System with software version VB10 about a potential malfunction of the motor controller.
In rare cases, the system processes incorrect internal values for the position of its ceiling tube stand. If in this situation the user then activates a longitudinal or transversal stand movement (by using the buttons at the tube head or the remote console), the system's controls may (depending on the values being processed) initiate a ""correction"" of the tube stand position, resulting in an unexpected movement.
Affected products
Ysio Radiography X-Ray System - Full Motorized
Lot or serial number
Plus de 10 numéros, contactez le fabricant.
Model or catalog number
10281013