Health product recall

Ysio Radiography X-Ray System - Full Motorized (2020-02-10)



Starting date:
February 10, 2020
Posting date:
February 28, 2020
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Laceration Hazard
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-72435


Last updated:

Affected Products

Ysio Radiography X-Ray System - Full Motorized

Reason

Siemens is notifying users of Ysio Fully Automated System with software version VB10 about a potential malfunction of the motor controller.

In rare cases, the system processes incorrect internal values for the position of its ceiling tube stand. If in this situation the user then activates a longitudinal or transversal stand movement (by using the buttons at the tube head or the remote console), the system's controls may (depending on the values being processed) initiate a ""correction"" of the tube stand position, resulting in an unexpected movement.

Affected products

Ysio Radiography X-Ray System - Full Motorized

Lot or serial number

Plus de 10 numéros, contactez le fabricant.

Model or catalog number

10281013

Companies
Manufacturer

Siemens AG

Wittelsbacherplatz 2

Muenchen

80333

GERMANY