Health product recall

Ysio Max (2018-08-08)

Starting date:
August 8, 2018
Posting date:
August 24, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-67594

Affected products

Ysio Max

Reason

Due to a problem with the steel cables inside the lifting column of the overhead tube, the steel cables may rupture. The probability of occurrence depends on the degree of usage of the system and is rated as extremely rare. In case the cables break, the tube arm may fall down unexpectedly during movement in vertical direction.

Affected products

Ysio Max

Lot or serial number
  • 24141
  • 24210
Model or catalog number

10762470

Companies
Manufacturer
Siemens Healthcare GmbH
Henkestr. 127
Erlangen
91052
GERMANY