XS and XT Hematology Analyzers (2016-11-18)
- Starting date:
- November 18, 2016
- Posting date:
- December 5, 2016
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-61304
Affected Products
A. XS-1000i, Hematology Analyzer
B. XT-2000i Hematology Analyzer
C. XN-10 Module
D. XN-11 Module
Reason
Sysmex America, Inc. (SAI) has become aware through customer complaints that the computer power supply on the information processing unit (IPU) used with the Sysmex XN-Series, XT-Series, and XS-Series hematology analyzers shut down, disabling the analyzer. The power supply has an over-voltage regulator to halt power in the presence of a power spike and prevent potential damage to the power supply and the associated device. The regulator in the power supplies within specific serial number range were found to be more sensitive than necessary to normal power fluctuations, causing the shutdown of the IPU during normal operation.
Affected products
A. XS-1000i, Hematology Analyzer
Lot or serial number
More than 10 numbers, contact manufacturer
Model or catalog number
XS-1000I
Companies
- Manufacturer
-
Sysmex Corporation
1-5-1 Wakinohama-kaigandori
Chuo-ku
651-0073
JAPAN
B. XT-2000i Hematology Analyzer
Lot or serial number
Plus de 10 numéros, contactez fabricant
Model or catalog number
XT-2000I
Companies
- Manufacturer
-
Sysmex Corporation
1-5-1 Wakinohama-kaigandori
Chuo-ku
651-0073
JAPAN
C. XN-10 Module
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
XN-10
Companies
- Manufacturer
-
Sysmex Corporation
1-5-1 Wakinohama-kaigandori
Chuo-ku
651-0073
JAPAN
D. XN-11 Module
Lot or serial number
More tha 10 numbers, contact manufacturer.
Model or catalog number
XN-11
Companies
- Manufacturer
-
Sysmex Corporation
1-5-1 Wakinohama-kaigandori
Chuo-ku
651-0073
JAPAN